PROGRAMAÇÃO
PARA AIRONE ®
| |||
| HDL DIRETO | |||
| Catálogo: | K.071 | Número de Testes: | 333 |
| TEST NAME |
HDL
DIRETO
|
| DECIMAL |
0
|
| MEASURE UNIT |
mg/dL
|
| REACTION TYPE |
END
POINT
|
| SAMPLE VOLUME (µl) |
3
|
| REAGENT VOLUME (µl) | |
| REAGENT 1 |
180
|
| REAGENT 2 |
60
|
| ABSORBANCE RANGE (mABS) | |
| MINIMUM |
0
|
| MAXIMUM |
400
|
| LINEARITY LIMIT |
200
|
| CONTAMINATING |
NO
|
| REAGENT BLANKING |
NO
|
| DIFFERENTIAL |
YES
|
| FILTERS (nm) | |
| FILTER 1 |
546
|
| FILTER 2 |
620
|
| TIMES (SEC) | |
| MIX 1 |
0.00
|
| MIX 2 |
0.20
|
| INCUBATION 1 |
300
|
| INCUBATION 2 |
300
|
| LAG PHASE |
3
|
| MEASURE |
1
|
| MEASUREMENT TYPE | |
| CALIBRATE |
STANDARD
|
| FACTOR |
0
|
Rev. Set/2009