PROGRAMAÇÃO
PARA ALCYON 300 ®
| |||
|
FERRO
SÉRICO AUTOMAÇÃO
| |||
|
Catálogo:
|
K.070
|
Número
de Testes:
|
400
|
| TEST |
FE
|
| Type |
End
Point
|
| Unit |
µg/dL
|
| Number of Decimals |
0
|
| Diluent |
99
Diluen
|
| Pre-washing |
No
|
| First Reagent Vol. (µL) |
200
|
| Second Reagent Vol. (µL) |
50
|
| Sample Volume (µL) |
50
|
| . |
.
|
| Lower Normal Value |
*
|
| Higer Normal Value |
*
|
| Systematic Dilution |
1
|
| Dilution for Urine |
1
|
| . |
.
|
| Low Reagent Blank |
0.0000
|
| High Reagent Blank |
0.1000
|
| Maximum Deviation |
0.0100
|
| . |
.
|
| Primary Wavelength |
550
nm
|
| High Wavelength |
.
|
| Low Wavelength |
.
|
| Systematic Serum Blank |
No
|
| . |
.
|
| T Tolerance Factor |
0.0100
|
| . |
.
|
| Antigen Excess Test (%) |
No
|
| . |
.
|
| Slope |
1.00
|
| Intercept |
0.0
|
| . |
.
|
| Reaction Delay |
03:00
|
Preparo dos Reagentes:
Reagente 1 : Tampão;
Reagente 2 : Misturar 2 mL do tampão com 0,5 mL do reagente de cor.
Rev. Out/2009