PROGRAMAÇÃO PARA COBAS MIRA PLUS®

TRANSFERRINA

Catálogo:

K.114-1

Número de Testes:

100

COBAS MIRA é marca registrada de seus proprietários.
Preparação e estabilidade do reagente de trabalho: utilizar as instruções de uso.

GENERAL CALCULATION
MEASUREMENT MODE: ABSORB SAMPLE LIMIT NO
REACTION MODE: D-R-S-SR1 REAC. DIRECTION INCREASE
CALIBRATION MODE: LOGIT / LOG5 CHECK OFF
REAGENT BLANK: NO BLANK ANTIGEN EXCESS NO
CLEANER: NO CONVERS. FACTOR 1.00000
WAVELENGHT: 340 nm OFFSET 0.00000
DECIMAL POSITION: 0 TEST RANGE LOW ON
UNIT: mg/dL HIGH ON
ANALYSIS NORM. RANGE LOW 170 mg/dL
SAMPLE DIL. NAME: H2O HIGH 340 mg/dL
FACTOR: NO NUMBER OF STEPS 1
TIME: NO CALC. STEP A ENPOINT
STD: MAIN INDIRECT READING FIRST: 4 LAST: 25
MAIN STD CONC. CALIB. . .
POS 7 . .

FACT. STD 1:1

2 : 2 . .
3 :2
4 : 2 CALIBRATION
5 :2
6 : 0 CALIB. INTERVAL ON REQUEST
7 : NO
8 : NO BLANK -
POST. DIL. FACTOR 2.00 REAG. RANGE LOW NO
CONC. FACTOR NO HIGH NO
SAMPLE CYCLE: 1 BLANK RANGE LOW NO
VOLUME: 2.0 µL HIGH NO
DILUENT NAME: H2O STANDARD POS *
VOLUME 10.0 µL 1: # 2: #
REAGENT: CYCLE: 1 3: # 4: #
VOLUME 220 µL 5: # 6: 0
START R1 CYCLE: 5 7: # NO 8: NO
VOLUME 20.0 µL REPLICATE SINGLE
DILUENT NAME: H2O DEVIATION NO
VOLUME 10 .0 µL CORRECTION STD NO
    CONTROL
. . CS1 POS: NO
. . CS2 POS: NO
. . CS3 POS: NO

* Parâmetros escolhidos pelo usuário

Ver Instruções de Uso

Rev. Out/2010