PROGRAMAÇÃO
PARA COBAS MIRA ®
| |||
| CK NAC – CRYSTAL | |||
| Catálogo: | K.010 | Número de Testes: | 344 |
| GENERAL | CALIBRATION | |||
| MEASUREMENT MODE: | ABSORB | CALIB. INTERVAL: | ON REQUEST | |
| REACTION MODE: | R-S | REAGENT BLANK | ||
| CALIBRATION MODE: | FACTOR | REAG. RANGE LOW: | 0.0900 A | |
| REAGENT BLANK: | REAG/DIL |
HIGH:
| 0.9000 A | |
| CLEANER: | BEFORE | BLANK RANGE LOW: | –0.0020 D A | |
| WAVELENGTH: | 340 nm |
HIGH:
| 0.0020 D A | |
| DECIMAL POSITION: | 0 | FACTOR: | 15696 | |
| UNIT: | U/L | CONTROL | ||
| ANALYSIS | CS1 POS: * Assign *** | Low*** High*** | ||
| SAMPLE DIL. NAME: | H2O | CS2 POS: * Assign *** | Low*** High*** | |
| POST DIL. FACTOR: | 3.00 | CS3 POS: * Assign *** | Low*** High*** | |
| CONC. FACTOR: | NO | |||
| SAMPLE CYCLE | 1 | |||
| VOL: | 3.0 µL | |||
| DIL: | 30.0 µL | |||
| REAGENT CYCLE: | 1 | |||
| VOL: | 145 µL | |||
| CALCULATIONS | ||||
| SAMPLE LIMIT: | 0.3000 D A | |||
| POINT: | T1 | |||
| REAC. DIRECTION: | INCREASE | |||
| CHECK: | ON | |||
| CONVERS. FACTOR: | 1.00000 | |||
| OFFSET: | 0.00000 | |||
| TEST RANGE LOW: | 0 U/L | |||
|
HIGH:
| 2000 U/L | |||
| NORM. RANGE LOW: | NO | |||
|
HIGH:
| NO | |||
| NUMBER OF STEPS: | 1 | |||
| CALC. STEP A | KINSEARCH | |||
| READINGS FIRST: | 5 | |||
| LAST: | 11 | |||
| REACTION LIMIT: | 0.6000 D A | |||
| POINT: | T1 |
Rev. nov/2004