PROGRAMAÇÃO PARA
LABMAX 240 ®
| |||
| PCR ULTRASENSÍVEL | |||
| Catálogo: |
K.079
| Número de Testes: | |
| ITEM NAME |
#
|
PCR
U
| |||
| DATA INFORMATION | |||||
| UNITS |
Mg/L
| ||||
| DECIMALS |
0
| ||||
| ANALYSIS | |||||
| TYPE |
END
| ||||
| MAIN W.LENGHT 1 |
505
| ||||
| SUB W.LENGHT 2 |
700
| ||||
| METHOD |
INIBIÇAO
SELETIVA
| ||||
| CORRELAÇÃO | |||||
|
Y=
|
SLOPE
|
X+
|
INTER
| ||
|
1
|
0
| ||||
| CALIBRATION | |||||
| TYPE |
LINEAR
| ||||
| STANDARD | |||||
|
#1
|
@
|
#4
| |||
|
#2
|
-
|
#5
| |||
|
#3
|
-
|
#6
| |||
| NORMAL RANGE | |||||
|
-
|
MALE
|
FEMALE
| |||
|
LOW
|
HIGH
|
LOW
|
HIGH
| ||
| SERUM |
0
|
1
|
0
|
1
| |
|
URINE
|
0
|
0
|
0
|
0
| |
|
PLASMA
|
0
|
0
|
0
|
0
| |
|
CSF
|
0
|
0
|
0
|
0
| |
|
DIALYSIS
|
0
|
0
|
0
|
0
| |
|
OTHER
|
0
|
0
|
0
|
0
| |
|
ASPIRATION
|
DATA
PROCESS READ
|
ABSORBANCE
LIMIT
|
||||
|
KIND
|
DOUBLE
|
--
|
START
|
END
|
LOW
|
-0.010
|
|
--
|
VOLUME
|
MAIN
|
53
|
54
|
HIGH
|
2500
|
|
SAMPLE
|
10
|
SUB
|
30
|
31
|
---
|
---
|
|
REAGENT1
VOL
|
100
|
-
|
-
|
-
|
-
|
-
|
|
REAGENT
2
|
100
|
-
|
-
|
-
|
-
|
-
|
|
THIRD
MIX
|
OFF
|
FACTOR
|
END
POINT LIMIT
|
---
|
||
|
R1
BLANK
|
WATER
BLANK
|
BLANK
CORRECTION
|
1.000
|
LINEAR
CHECK(%)
|
---
|
|
|
-
|
||||||
|
MONITOR
|
PROZONE
CHECK
|
|||||
|
0
LEVEL POINT
|
1
|
-
|
START
|
END
|
LIMIT(%)
|
-
|
|
SPAN
|
3
|
FIRST
|
-
|
-
|
-
|
-
|
|
-
|
-
|
SECOND
|
-
|
-
|
-
|
LOW
|
|
-
|
-
|
THIRD
|
-
|
-
|
-
|
LOW
|
|
AUTO
RETURN SW
|
-
|
AUTO
RETURN CONDITION (ABSORBANCE)
|
|||
|
-
|
OFF
|
ABSORBANCE
RANGE
|
-
|
||
|
--
|
LOWER
|
OFF
|
|||
|
AUTO
RETURN RANGE
|
HIGHER
|
OFF
|
|||
|
-
-
|
OFF
|
OFF
|
-
|
-
|
|
|
LOWER
|
HIGHER
|
PROZONE
RANGE
|
OFF
|
||
|
SERUM
|
0
|
20
|
-.
-.
-
-
|
||
|
URINE
|
-
|
-
|
|||
|
PLASMA
|
-
|
-
|
|||
|
CSF
|
-
|
-
|
|||
|
DIALYSIS
|
-
|
-
|
|||
|
OTHER
|
-
|
-
|
|||
Rev. Maio/2009