Serviço
de Assessoria Científica - 0800-315454
|
PROGRAMAÇÃO
PARA SELECTRA II e E, XL e FLEXOR ®
| ||||
|
CK
NAC - CRYSTAL
| ||||
|
Catálogo:
|
K.010
|
Número
de Testes:
|
178
| |
Preparação
e estabilidade do reagente de trabalho: utilizar as instruções de uso.
TEST PARAMETER MONO/DUAL MODE – 37 ºC
| Name |
CK-NAC
|
Prozone check |
No
|
| Abbr.Name |
CK
|
Ref. Male low |
*
|
| Mode |
Kinetic
|
Ref. Male high |
*
|
| Wavelength |
340
nm
|
Ref. Female high |
*
|
| Units |
U/L
|
Ref. Ped low |
*
|
| Decimals |
0
|
Ref. Ped high |
*
|
| Low Conc. |
0.00
U/L
|
Control 1 |
BIOCONTROL
|
| High Conc. |
2000
U/L
|
. |
*
|
| Calibrator Name |
*
|
Control 2 |
*
|
| Repeat |
*
|
. |
*
|
| Number |
*
|
Control 3 |
*
|
| Concentration |
*
|
. |
*
|
| Interval |
*
|
Correlat factor |
1.000
|
| Cut off |
*
|
Correlat offset |
0.000
U/L
|
| DUAL MODE |
.
|
. |
.
|
| Name |
CK-NAC
|
Linearyt Limit |
20
%
|
| Sample blank |
No
|
Low Absorbance |
-0.100
Abs
|
| R1 bottle |
25 mL
|
High Absorbance |
3.000
Abs
|
| Normal volume |
280 µL
|
R.Abs.L.Limit |
-0.100
Abs
|
| Rerun volume |
281
µL
|
R.Abs.H.Limit |
3.000
Abs
|
| Sample |
-
|
R.Abs. Deviation |
3.000
Abs
|
| Normal volume |
6
µL
|
Reagent Blank |
Yes
(#)
|
| Rerun volume |
3
µL
|
. |
*
|
| R2 bottle |
5
mL
|
. |
*
|
| Normal volume |
0.0
µL
|
. |
*
|
| Rerun volume |
0.0
µL
|
Factor |
8095
|
| Predilution |
No
|
. |
*
|
| Sloope Blank |
No
|
. |
*
|
| Delay,min. time |
130.177
sec
|
. |
*
|
| MONO MODE |
.
|
. |
.
|
| Name |
CK-NAC
|
Linearyt Limit |
10
% Sample
|
| Sample blank |
No
|
Low Absorbance |
0.000
Abs
|
| R1 bottle |
25
mL
|
High Absorbance |
3.000
Abs
|
| Normal volume |
250
µL
|
R.Abs.L.Limit |
0.000
Abs
|
| Rerun volume |
252 µL
|
R.Abs.H.Limit |
0.800
Abs
|
| Sample |
-
|
R.Abs. Deviation |
3.000
Abs
|
| Normal volume |
5
µL
|
Reagent Blank |
Yes
(#)
|
| Rerun volume |
3.0
µL
|
Cal Low Limit | - |
| - |
-
|
Cal High Limit | -- |
| Delay,min. time |
129.293
sec
|
Factor |
8095
|
*Selecionado
pelo usuário. .....#Calculado pelo analisador
Um soro
controle recem reconstituído deve ser utilizado para ajustar
o fator.
Isso acontece devido a variação da potência da lâmpada
e a temperatura de incubação.
Ver
Instruções de Uso
Rev. Mar/2010