PROGRAMAÇÃO PARA
ABBOTT VP ®
| |||
| GAMA GT LÍQUIDO ESTÁVEL | |||
| Catálogo: | K.080-2 | Número de Testes: | 240 |
| TEST NUMBER |
**
|
xx
|
| TEST NAME |
GGTP
|
12
|
| TEMPERATURE ºC |
37
|
37
|
| FILTER VALUE |
415/450
|
50
|
| UNIT OF MEASUREMENT |
U/L
|
11
|
| DILUTION RATIO |
1:51
|
51
|
| REV TIME |
2
|
2
|
| AUX DISP |
NÃO
|
N
|
| FRR ? |
SIM
|
Y
|
| REACTION UP ? |
SIM
|
Y
|
| STANDARDS ? |
NÃO
|
N
|
| RGT BLK |
SIM
|
Y
|
| ASSAY FACTOR 25500 | ||
| END POINT |
NÃO
|
N
|
| BEGIN PRINT REVOLUTION |
3
|
3
|
| Nº PRINT REVOLUTION |
1
|
1
|
| INITIAL ABS |
0,64
|
0.64
|
| UP LIMIT |
SIM
|
Y
|
| SUBS DEP |
1,2
|
1.2
|
| REAG DGR |
20
|
20
|
Rev. Fer/2010