Serviço
de Assessoria Científica - 0800-315454
|
PROGRAMAÇÃO
PARA ALCYON 300 ®
| |||
|
CÁLCIO
ARSENAZO III
| |||
|
Catálogo:
|
T.009
|
Número
de Testes:
|
333
|
Preparação e
estabilidade do reagente de trabalho: utilizar as instruções de uso.
| TEST |
CALCIO
|
| Type |
End
Pt.
|
| Unit |
mg/dL
|
| Number of Decimals |
2
|
| Diluent |
99
Diluen
|
| Pre-washing |
No
|
| First Reagent Vol. (µL) |
300
|
| Second Reagent Vol. (µL) |
0
|
| Sample Volume (µL) |
3.0
|
| . |
.
|
| Lower Normal Value |
8.8
|
| Higer Normal Value |
11.0
|
| . |
.
|
| Systematic Dilution |
1
|
| Dilution for Urine |
1
|
| . |
.
|
| Low Reagent Blank |
0.0000
|
| High Reagent Blank |
3.5000
|
| Maximum Deviation |
0.0150
|
| . |
.
|
| Primary Wavelength |
600
nm
|
| High Wavelength |
..
|
| Low Wavelength |
..
|
| Systematic Serum Blank |
No
|
| . |
.
|
| T Tolerance Factor |
0.0150
|
| . |
.
|
| Antigen Excess Test (%) |
No
|
| . |
.
|
| Slope |
1.00
|
| Intercept |
0.0
|
| . |
.
|
| Reaction Delay |
01:00
|
Rev. nov/2004