PROGRAMAÇÃO
PARA ALCYON 300 ®
| |||
|
GAMA
GT LÍQUIDO ESTÁVEL
| |||
|
Catálogo:
|
K.080-2
|
Número
de Testes:
|
300
|
Preparação e estabilidade
do reagente de trabalho: utilizar as instruções de uso.
| TEST |
GGT
|
| Type |
Kin.
0
|
| Unit |
U/L
|
| Number of Decimals |
0
|
| Diluent |
99
Diluen
|
| Pre-washing |
No
|
| First Reagent Vol. (µl) |
200
|
| Second Reagent Vol. (µl) |
0
|
| Sample Volume (µl) |
10.0
|
| . |
.
|
| Lower Normal Value |
*
|
| Higer Normal Value |
*
|
| Dilution Factor |
10
|
| Systematic Dilution |
1
|
| Dilution for Urine |
1
|
| . |
.
|
| Low Reagent Blank |
0.0000
|
| High Reagent Blank |
0.6000
|
| Maximum Deviation |
0.0100
|
| . |
.
|
| Primary Wavelength |
405
nm
|
| High Wavelength |
.
|
| Low Wavelength |
.
|
| Systematic Serum Blank |
No
|
| . |
.
|
| High Limit |
0.0100
|
| Depletion Limit |
0.1560
|
| C Compensation Factor |
0.0001
|
| . |
.
|
| Slope |
1.00
|
| Intercept |
0.0
|
| . |
.
|
| Reaction Delay |
01:00
|
Para calcular o fator de calibração dividir 44400 pelo coeficiente de extinção molar do PNA informado no aparelho.