PROGRAMAÇÃO
PARA ALCYON 300 ®
| |||
|
GLICOSE
NOMOREAGENTE
| |||
|
Catálogo:
|
K.082-2
|
Número
de Testes:
|
1666
|
Preparação e estabilidade do reagente de trabalho: utilizar as instruções de uso.
| TEST |
GLIC
|
| Type |
Kin.
1
|
| Unit |
mg/dL
|
| Number of Decimals |
0
|
| Diluent |
99
Diluen
|
| Pre-washing |
No
|
| First Reagent Vol. (µL) |
300
|
| Second Reagent Vol. (µL) |
0
|
| Sample Volume (µL) |
3.0
|
| . |
.
|
| Lower Normal Value |
*
|
| Higer Normal Value |
*
|
| Systematic Dilution |
1
|
| Dilution for Urine |
1
|
| . |
.
|
| Low Reagent Blank |
0.0000
|
| High Reagent Blank |
0.1500
|
| Maximum Deviation |
0.0100
|
| . |
.
|
| Primary Wavelength |
500
nm
|
| High Wavelength |
.
|
| Low Wavelength |
.
|
| Systematic Serum Blank |
No
|
| . |
.
|
| Depletion Limit |
3.5000
|
| T Tolerance Factor |
1.0000
|
| C Compensation Factor |
1.0000
|
| . |
.
|
| Slope |
1.00
|
| Intercept |
0.0
|
| . |
.
|
| Reaction Delay |
00:24
|
| Incubation |
01:00
|
| Total |
01:24
|