PROGRAMAÇÃO PARA COBAS MIRA PLUS®

COMPLEMENTO C3

Catálogo:

K.076-1

Número de Testes:

166

COBAS MIRA é marca registrada de seus proprietários.

GENERAL CALCULATION
MEASUREMENT MODE: ABSORB SAMPLE LIMIT NO
REACTION MODE: D-R-S-SR1 REAC. DIRECTION INCREASE
CALIBRATION MODE: LOGIT / LOG5 CHECK ON
REAGENT BLANK: REAG/DIL ANTIGEN EXCESS NO
CLEANER: NO CONVERS. FACTOR 1.00000
WAVELENGHT: 340 nm OFFSET 0.00000
DECIMAL POSITION: 1 TEST RANGE LOW ON
UNIT: mg/dL HIGH ON
ANALYSIS NORM. RANGE LOW 78
SAMPLE DIL. NAME: H2O HIGH 135
FACTOR: 10.00 NUMBER OF STEPS 1
TIME: NO CALC. STEP A ENPOINT
STD: MAIN INDIRECT READING FIRST: 4 LAST:20
MAIN STD CONC. CAL. x10 - -
POS ** CALIBRATION

FACT. STD ...1 : 6

2 : 2 CALIB. INTERVAL ON REQUEST
3 : 2
4 : 2 BLANK -
5 : 2
6 : 0 REAG. RANGE LOW NO
7 : NO
8 : NO HIGH NO
POST. DIL. FACTOR 2.00 BLANK RANGE LOW NO
CONC. FACTOR NO HIGH NO
SAMPLE CYCLE: 1 STANDARDS -
VOLUME: 15.0 µL 1: # 2: #
DILUENT NAME: H2O 3: # 4: #
VOLUME 10.0 µL 5: # 6: 0
REAGENT: CYCLE: 1 7: # NO 8: NO
VOLUME 150 µL REPLICATE SINGLE
START R1 CYCLE 5 DEVIATION NO
VOLUME 12 µL CORRECTION STD NO
DILUENT NAME H2O - -
VOLUME 10 µL CONTROL
    CS1 POS: *
    CS2 POS: *
    CS3 POS: *
- - - -

* Parâmetros escolhidos pelo usuário

Ver Instruções de Uso

Rev. Jan/2007